Autoclave Calibration and Validation Q&A

Got a Question? Request a Callback

Lab tech using autoclave

Why do I need an autoclave calibration?

The process of calibration allows us to confirm the accuracy of the machine and its associated displays and recorders by comparing against a calibrated reference standard.


Why do I need an autoclave validation?

The process of validation allows us to confirm that an entire load has achieved the required temperature for the required time. A validation also includes a calibration as part of the process.
Validation is recommended: -

  • For new or previously unvalidated loads
  • Following changes to previously validated loads i.e. load content, temperature or time
  • After major repairs to an autoclave


What equipment will the engineer bring?

The engineer will bring along all the equipment needed, including a chart recorder, probes, and a laptop. The chart recorder is placed close to the autoclave and the probes are fed into the chamber.


How should I prepare for the visit?

For calibration, no specific preparation is required.

For validation, a representative load is required – not necessarily media as water could be used.


How long will the validation process take?

The chart recorder takes about an hour to ‘warm up’ and in the meantime, the engineer will start to prepare your certificate. A typical calibration takes approximately 4 hours. A validation will take longer and will vary depending on the load content.


Autoclave inner

When will I be able to see the data?

The data is shown, in real time, on the chart recorder screen and the engineer will be able to explain the results during the visit.


When will I get the certificate?

Your certificate is prepared during the visit and will be sent to the office to be checked by an approved signatory before being issued. Typical turnaround time is less than 5 working days.


For more information on our Calibration and Validation services, please visit UKAS Calibration / Validation | Don Whitley Scientific (dwscientific.com) or call our service team on 01274 595728 or email service@dwscientific.co.uk

UKAS banner


Chat

Choose a Country

If you are based in Germany or Australia, please select the website of our subsidiary company there. A different selection of products is available in these countries. Otherwise, please select ‘English’ and you will be able to find your local distributor in the ‘Overseas Distributors’ section.

Share this

Please select your preferred language from the list below

Login

Don't have an account? Click Here

Forgot your password? Click Here